A unique blend of high-performance hot-start MyTaq™ HS DNA Polymerase and a proprietary proofreading enzyme, specifically developed for long PCR applications
RANGER Mix is an easy-to-use, high-performance enzyme blend, specifically developed to amplify fragments up to 25 kb in length. RANGER contains a unique combination of a highly-efficient DNA polymerases and novel buffer system that deliver the improved efficiency necessary for reliable amplification of longer amplicons.
RANGER is an antibody-mediated hot-start enzyme blend that eliminates non-specific amplification during reaction set-up. The inactivated enzymes do not possess polymerase activity, thereby preventing the non-specific amplification, such as primer-dimer formation, that often hinder long PCR reactions from the start.
RANGER Mix is an all-in-one mastermix containing dNTPs, MgCl2 and enhancers at optimal concentrations, minimizing the requirement for PCR optimization and driving greater sensitivity. RANGER Mix also improves reproducibility and minimizes the requirement for assay repeats, by reducing the number of pipetting steps, and therefore the risk of manual error, during reaction set-up.
Reagent |
500 Reactions |
RANGER Mix, 2x |
10 x 1.25 mL |
All components should be stored at -20°C upon receipt for optimum stability. Repeated freeze/thaw cycles should be avoided. When stored under the recommended conditions and handled correctly, full activity of the reagents is retained until the expiry date indicated on the outer box label.
Components may also be stored at +4°C if required, although storage at -20°C is recommended. When stored at +4°C, reagents will remain stable for a period of 2 weeks from date of receipt.
Shipped on Dry Ice or Blue Ice
Bioline's mixes contain magnesium chloride at the following concentrations (please see table below). An additional tube of 50 mM MgCl2 solution is supplied with each mix to facilitate further optimization if required.
Bioline Mix | Final Magnesium Concentration |
ACCUZYME Mix | 2.0 mM. |
BioMix / BioMix Red | 2.5 mM. |
ImmoMix / ImmoMix Red | 3.0 mM. |
BIO-X-ACT Short Mix | 2.0 mM. |
MangoMix | 2.5 mM. |
MyTaq | 3.0 mM. |
RANGER | 1.5 mM. |
Observation | Recommended Solution(s) |
No or low PCR yield | Enzyme concentration too low – increase the amount of enzyme in 0.5 U increments. |
Primers degraded – check quality and age of the primers. | |
Magnesium concentration too low – increase concentration in 0.25 mM increments with a starting concentration of 1.75 mM. | |
Primer concentration not optimized. Titrate primer concentration (0.3-1 µM); ensuring that both primers have the same concentration. | |
Template concentration too low – Increase concentration of template. | |
Perform a positive control to ensure that the enzyme, dNTPs and buffers are not degraded and/or contaminated. | |
Multiple Bands | Primer annealing temperature too low. Increase annealing temperature. Primer annealing should be at least 5°C below the calculated Tm of primers. |
Prepare master mixes on ice or use a heat-activated polymerase. | |
For problems with low specificity. Try adding 3% DMSO (not supplied) to improve specificity. | |
Smearing or artifacts | Template concentration too high. Prepare serial dilutions of template. |
Too many cycles. Reduce the cycle number by 3-5 to remove non-specific bands. | |
Enzyme concentration too high - decrease the amount of enzyme in 0.5 U increments. | |
Extension time too long. Reduce extension time in 0.5-1 minute increments. |
Please click here in order to request your sample. You will receive an email confirmation within two business days with delivery details.